Wear mode and wear mechanism of retrieved acetabular cups

  • Kengo Yamamoto
  • , Atsuhiro Imakiire
  • , Toshinori Masaoka
  • , Takaaki Shishido
  • , Tatsuro Mizoue
  • , Ian C. Clarke
  • , Hiromu Shoji
  • , Keiichi Kawanabe
  • , Jiro Tamura

Research output: Contribution to journal › Article › peer-review

Abstract

Nineteen ultra-high-molecular-weighted polyethylene (UHMWPE) cups were retrieved at revision surgery. The implant period was on average 9.9 (0.25-20.7) years. The cups were classified into the high-wear group (wear rate above 140 mm3/year), intermediate-wear group (wear rate 80-140 mm 3/year) and low-wear group (wear rate below 80 mm3/year). The wear rate was measured by a fluid displacement method. The cups were studied using scanning electron microscopy. Ripples mainly appeared on the cups in the low-wear group, and nodules and fibrils mainly appeared in the high-wear group. Folding was observed in all cups, but folding with numerous fibrils was conspicuous in the high-wear group. The cup retrieved at 3 months already exhibited folding. From these results, the wear response appeared to be a three-step process: (1) foldings are generated, (2) ripples are formed on the surface, and (3) fibrils are produced from the surface and delaminated, leading to the formation of wear debris.

Original languageEnglish
Pages (from-to)286-290
Number of pages5
JournalInternational Orthopaedics
Volume27
Issue number5
DOIs
StatePublished - Jun 21 2003