Validation of a 3D CT method for measurement of linear wear of acetabular cups: A hip simulator study

  • Anneli Jedenmalm
  • , Fritjof Nilsson
  • , Marilyn E. Noz
  • , Douglas D. Green
  • , Ulf W. Gedde
  • , Ian C. Clarke
  • , Andreas Stark
  • , Gerald Q. Maguire
  • , Michael P. Zeleznik
  • , Henrik Olivecrona

Research output: Contribution to journal › Article › peer-review

Abstract

Background We evaluated the accuracy and repeatability of a 3D method for polyethylene acetabular cup wear measurements using computed tomography (CT). We propose that the method be used for clinical in vivo assessment of wear in acetabular cups. Material and methods Ultra-high molecular weight polyethylene cups with a titanium mesh molded on the outside were subjected to wear using a hip simulator. Before and after wear, they were (1) imaged with a CT scanner using a phantom model device, (2) measured using a coordinate measurement machine (CMM), and (3) weighed. CMM was used as the reference method for measurement of femoral head penetration into the cup and for comparison with CT, and gravimetric measurements were used as a reference for both CT and CMM. Femoral head penetration and wear vector angle were studied. The head diameters were also measured with both CMM and CT. The repeatability of the method proposed was evaluated with two repeated measurements using different positions of the phantom in the CT scanner. Results The accuracy of the 3D CT method for evaluation of linear wear was 0.51 mm and the repeatability was 0.39 mm. Repeatability for wear vector angle was 17°. Interpretation This study of metal-meshed hip-simulated acetabular cups shows that CT has the capacity for reliable measurement of linear wear of acetabular cups at a clinically relevant level of accuracy. © 2011 Nordic Orthopaedic Federation.
Original languageAmerican English
Pages (from-to)35-41
Number of pages7
JournalActa Orthopaedica
Volume82
Issue number1
DOIs
StatePublished - Feb 2011

ASJC Scopus Subject Areas

  • Surgery
  • Orthopedics and Sports Medicine

Keywords

  • Polyethylene
  • Reproducibility of Results
  • Stress, Mechanical
  • Acetabulum
  • Tomography, X-Ray Computed
  • Hip Prosthesis/adverse effects
  • Prosthesis Design
  • Biomechanical Phenomena
  • Prosthesis Failure
  • Models, Biological
  • Surface Properties
  • Materials Testing/methods
  • Titanium
  • Biocompatible Materials
  • Imaging, Three-Dimensional
  • Equipment Failure Analysis

Disciplines