TU-FG-BRB-04: A New Optimization Method for Pre-Treatment Patient-Specific Stopping-Power by Combining Proton Radiography and X-Ray CT

C Collins-Fekete, Reinhard W Schulte, L Beaulieu, J Seco

Research output: Contribution to journalMeeting abstractpeer-review

Abstract

Purpose: The relative stopping power (RSP) uncertainty is the largest contributor to the range uncertainty in proton therapy. The purpose of this work is to develop a robust and systematic method that yields accurate patient specific RSPs by combining pre-treatment X-ray CT and daily proton radiography. Methods: The method is formulated as a penalized least squares optimization (PLSO) problem min(, Ax-B, ). The matrix A represents the cumulative path-length crossed in each material computed by calculating proton trajectories through the X-ray CT. The material RSPs are denoted by x and B is the pRad, expressed as water equivalent thickness. The equation is solved using a convex-conic optimizer. Geant4 simulations were made to assess the feasibility of the method. RSP extracted from the Geant4 materials were used as a reference and the clinical HU-RSP curve as a comparison. The PLSO was first tested on a Gammex RMI-467 phantom. Then, anthropomorphic phantoms of the head, pelvis and lung were studied and resulting RSPs were evaluated. A pencil beam was generated in each phantom to evaluate the proton range accuracy achievable by using the optimized RSPs. Finally, experimental data of a pediatric head phantom (CIRS) were acquired using a recently completed experimental pCT scanner. Results: Numerical simulations showed precise RSP (
Original languageAmerican English
Pages (from-to)3756-3757
Number of pages2
JournalMedical Physics
Volume43
Issue number6Part35
DOIs
StatePublished - Jun 1 2016

Disciplines

  • Physics
  • Other Medical Specialties

Cite this