TY - JOUR
T1 - Silicon nitride bearings for total joint arthroplasty
AU - McEntire, Bryan J.
AU - Lakshminarayanan, Ramaswamy
AU - Ray, Darin A.
AU - Clarke, Ian C.
AU - Puppulin, Leonardo
AU - Pezzotti, Giuseppe
N1 - Publisher Copyright:
© 2016 by the authors; licensee MDPI, Basel, Switzerland.
PY - 2016/12/1
Y1 - 2016/12/1
N2 - The articulation performance of silicon nitride against conventional and highly cross-linked polyethylene, as well as for self-mated silicon nitride bearings, was examined in a series of standard hip simulation studies. Wear rates for polyethylene liners against silicon nitride femoral heads were consistent with reported literature, although higher than cobalt chromium controls. Excessive protein precipitation was a confounding factor in interpretation of the wear data. Post wear-test Raman spectroscopy of the cross-linked polyethylene liners showed no oxidative degradation. Wear of self-mated silicon nitride was found to be essentially zero and indistinguishable from alumina controls using continuously orbital hip simulation for up to three million cycles. However, introduction of an alternative loading profile from three to five million cycles, including a stop-dwell-start sequence, significantly increased wear for two of six silicon nitride couples. This behavior is associated with formation and disruption of a gelatinous silicic acid tribochemical film, and is consistent with a recurrent transition from fluid-film to boundary lubrication. Overall, these results suggest that silicon nitride articulation against dissimilar counterface surfaces (e.g., highly cross-linked polyethylene) is preferred.
AB - The articulation performance of silicon nitride against conventional and highly cross-linked polyethylene, as well as for self-mated silicon nitride bearings, was examined in a series of standard hip simulation studies. Wear rates for polyethylene liners against silicon nitride femoral heads were consistent with reported literature, although higher than cobalt chromium controls. Excessive protein precipitation was a confounding factor in interpretation of the wear data. Post wear-test Raman spectroscopy of the cross-linked polyethylene liners showed no oxidative degradation. Wear of self-mated silicon nitride was found to be essentially zero and indistinguishable from alumina controls using continuously orbital hip simulation for up to three million cycles. However, introduction of an alternative loading profile from three to five million cycles, including a stop-dwell-start sequence, significantly increased wear for two of six silicon nitride couples. This behavior is associated with formation and disruption of a gelatinous silicic acid tribochemical film, and is consistent with a recurrent transition from fluid-film to boundary lubrication. Overall, these results suggest that silicon nitride articulation against dissimilar counterface surfaces (e.g., highly cross-linked polyethylene) is preferred.
KW - Friction
KW - Hip simulation
KW - Polyethylene
KW - Silicon nitride
KW - Tribochemical film
KW - Wear
UR - http://www.scopus.com/inward/record.url?scp=85046280712&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85046280712&partnerID=8YFLogxK
U2 - 10.3390/lubricants4040035
DO - 10.3390/lubricants4040035
M3 - Article
SN - 2075-4442
VL - 4
JO - Lubricants
JF - Lubricants
IS - 4
M1 - 35
ER -