TY - JOUR
T1 - Precision and consistency of the passive leg raising maneuver for determining fluid responsiveness with bioreactance noninvasive cardiac output monitoring in critically ill patients and healthy volunteers
AU - Chopra, Sahil
AU - Thompson, Jordan
AU - Shahangian, Shahab
AU - Thapamagar, Suman
AU - Moretta, Dafne
AU - Gasho, Chris
AU - Cohen, Avi
AU - Bryant Nguyen, H.
N1 - Publisher Copyright:
© 2019 Chopra et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
PY - 2019/9/1
Y1 - 2019/9/1
N2 - OBJECTIVE: The passive leg raising (PLR) maneuver has become standard practice in fluid resuscitation. We aim to investigate the precision and consistency of the PLR for determining fluid responsiveness in critically ill patients and healthy volunteers using bioreactance non-invasive cardiac output monitoring (NiCOM™, Cheetah Medical, Inc., Newton Center, Massachusetts, USA).METHODS: This study is prospective, single-center, observational cohort with repeated measures in critically ill patients admitted to the medical intensive care unit and healthy volunteers at a tertiary academic medical center. Three cycles of PLR were performed, each at 20-30 minutes apart. Fluid responsiveness was defined as a change in stroke volume index (ΔSVI) > 10% with each PLR as determined by NiCOM™. Precision was the variability in ΔSVI after the 3 PLR's, and determined by range, average deviation and standard deviation. Consistency was the same fluid responsiveness determination of "Yes" (ΔSVI > 10%) or "No" (ΔSVI ≤ 10%) for all 3 PLR's.RESULTS: Seventy-five patients and 25 volunteers were enrolled. In patients, the precision was range of 17.2±13.3%, average deviation 6.5±4.0% and standard deviation 9.0±5.2%; and for volunteers, 17.4±10.3%, 6.6±3.8% and 9.0±6.7%, respectively. There was no statistical difference in the precision measurements between patients and volunteers. Forty-nine (65.3%) patients vs. twenty-four (96.0%) volunteers had consistent results, p < 0.01. Among those with consistent results, twenty-four (49.0%) patients and 24 (100%) volunteers were fluid responsive.CONCLUSIONS: The precision and consistency of determining ΔSVI with NiCOM™ after PLR may have clinical implication if ΔSVI > 10% is the absolute cutoff to determine fluid responsiveness.
AB - OBJECTIVE: The passive leg raising (PLR) maneuver has become standard practice in fluid resuscitation. We aim to investigate the precision and consistency of the PLR for determining fluid responsiveness in critically ill patients and healthy volunteers using bioreactance non-invasive cardiac output monitoring (NiCOM™, Cheetah Medical, Inc., Newton Center, Massachusetts, USA).METHODS: This study is prospective, single-center, observational cohort with repeated measures in critically ill patients admitted to the medical intensive care unit and healthy volunteers at a tertiary academic medical center. Three cycles of PLR were performed, each at 20-30 minutes apart. Fluid responsiveness was defined as a change in stroke volume index (ΔSVI) > 10% with each PLR as determined by NiCOM™. Precision was the variability in ΔSVI after the 3 PLR's, and determined by range, average deviation and standard deviation. Consistency was the same fluid responsiveness determination of "Yes" (ΔSVI > 10%) or "No" (ΔSVI ≤ 10%) for all 3 PLR's.RESULTS: Seventy-five patients and 25 volunteers were enrolled. In patients, the precision was range of 17.2±13.3%, average deviation 6.5±4.0% and standard deviation 9.0±5.2%; and for volunteers, 17.4±10.3%, 6.6±3.8% and 9.0±6.7%, respectively. There was no statistical difference in the precision measurements between patients and volunteers. Forty-nine (65.3%) patients vs. twenty-four (96.0%) volunteers had consistent results, p < 0.01. Among those with consistent results, twenty-four (49.0%) patients and 24 (100%) volunteers were fluid responsive.CONCLUSIONS: The precision and consistency of determining ΔSVI with NiCOM™ after PLR may have clinical implication if ΔSVI > 10% is the absolute cutoff to determine fluid responsiveness.
KW - Fluid Therapy
KW - Patient Positioning/methods
KW - Intensive Care Units
KW - Prospective Studies
KW - Humans
KW - Middle Aged
KW - Male
KW - Stroke Volume/physiology
KW - Treatment Outcome
KW - Critical Illness/therapy
KW - Healthy Volunteers
KW - Monitoring, Physiologic/methods
KW - Feasibility Studies
KW - Leg
KW - Young Adult
KW - Adult
KW - Female
KW - Aged
UR - https://www.scopus.com/pages/publications/85072693333
UR - https://www.scopus.com/pages/publications/85072693333#tab=citedBy
UR - https://www.mendeley.com/catalogue/d2571c51-1074-37d5-8a57-9c6ce1e76116/
U2 - 10.1371/journal.pone.0222956
DO - 10.1371/journal.pone.0222956
M3 - Article
C2 - 31560711
SN - 1932-6203
VL - 14
SP - e0222956
JO - PLOS One
JF - PLOS One
IS - 9
M1 - e0222956
ER -