Knotted ureteral stent: A minimally invasive technique for removal

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose: We describe a useful technique for untying a knotted ureteral catheter. Materials and Methods: An Amplatz super stiff guide wire was advanced through the lumen of the stent and with uniform force the knot was untied. Results: The knot was untied and the stent was removed. A nephrostomy tube or other more invasive techniques for removal were avoided. Conclusions: An Amplatz super stiff guide wire passed retrograde via the ureteral stent lumen should be considered as a valid alternative to percutaneous removal of a knotted ureteral stent.

Original languageEnglish
Pages (from-to)2065-2066
Number of pages2
JournalThe Journal of urology
Volume159
Issue number6
DOIs
StatePublished - Jun 1998

ASJC Scopus Subject Areas

  • Urology

Keywords

  • Adverse effects
  • Stents
  • Ureter

Cite this