Abstract
We develop a standardized, fully automated, quantification system for liver fibrosis assessment using second harmonic generation microscopy and a morphology-based quantification algorithm. Liver fibrosis is associated with an abnormal increase in collagen as a result of chronic liver diseases. Histopathological scoring is the most commonly used method for liver fibrosis assessment, where a liver biopsy is stained and scored by experienced pathologists. Due to the intrinsic limited sensitivity and operator-dependent variations, there exist high inter- and intraobserver discrepancies. We validate our quantification system, Fibro-C-Index, with a comprehensive animal study and demonstrate its potential application in clinical diagnosis to reduce inter- and intraobserver discrepancies.
Original language | English |
---|---|
Article number | 044013 |
Journal | Journal of Biomedical Optics |
Volume | 14 |
Issue number | 4 |
DOIs | |
State | Published - 2009 |
ASJC Scopus Subject Areas
- Electronic, Optical and Magnetic Materials
- Biomaterials
- Atomic and Molecular Physics, and Optics
- Biomedical Engineering
Keywords
- bioimaging
- diagnosis
- liver fibrosis
- medical optics
- second harmonic generation
- two photon