Skip to main navigation Skip to search Skip to main content

Experimental validation of track structure models

    Research output: Chapter in Book/Report/Conference proceeding › Conference contribution

    Abstract

    A tracking ion counting nanodosimeter was employed to acquire spatial ionization patterns produced by charged particles in propane gas at 1.3 mbar. Data were taken at the James M. Slater MD Proton Treatment and Research Center with 250 MeV, 17 MeV, 5 MeV and 1.5 MeV proton beams, 4.8 MeV alpha particles and electrons from a Sr-90/Y-90 source. For each particle type, measurable quantities used for track structure reconstruction included the number of ionizations and their location within a wall-less, cylindrical sensitive volume measured with a resolution of about 5 tissue-equivalent nanometers, and primary particle coordinates. Measured ionization frequency distributions as a function of distance from particle track were compared with results of a dedicated Monte Carlo track structure code. © 2009 IEEE.
    Original languageEnglish
    Title of host publicationIEEE Transactions on Nuclear Science
    PublisherIEEE
    Pages2859-2863
    Number of pages5
    Volume56
    Edition5
    DOIs
    StatePublished - Oct 6 2009

    Publication series

    NameIEEE Transactions on Nuclear Science
    ISSN (Print)0018-9499
    ISSN (Electronic)1558-1578

    ASJC Scopus Subject Areas

    • Nuclear and High Energy Physics
    • Nuclear Energy and Engineering
    • Electrical and Electronic Engineering

    Keywords

    • Ionization
    • Monte Carlo methods
    • Particle track structure
    • Particle tracking
    • Radiation detectors
    • Radiation effects