Evaluation of the powder-formulated enzyme multiplied immunoassay technique quantitative single test for gentamicin

  • Lee S. Berk
  • , Nemi Imperio
  • , William C. Eby

Research output: Contribution to journalArticlepeer-review

Abstract

The usefulness of the enzyme multiplied immunoassay quantitative single test (EMIT QST) gentamicin assay was assessed for gentamicin analysis in patient sera. The EMIT QST reagents are in powder form in a single, premeasured vial and are run on a thermoregulated sample processor that controls mixing and timing steps. The results of the clinical evaluation showed that the standard curve was stable throughout a 26-day study period. Within-run precision on 20 replicates at 4.0 μg/ml yielded a coefficient of variation (CV) of 5.6%; between-run precision on 66 analyses at 6.0 μ/ml over a 152-day period yielded a CV of 4.0%. Mean recovery through the range of the standard curve with 10 spiked patient samples was 102%. Comparative analysis with radioimmunoassay of 95 patient samples showed a correlation of 0.97, with y = 0.93 - 0.03. It was concluded that the EMIT QST gentamicin assay is an appropriate, rapid methodology for patient gentamicin analysis.

Original languageEnglish
Pages (from-to)111-114
Number of pages4
JournalTherapeutic Drug Monitoring
Volume8
Issue number1
DOIs
StatePublished - Mar 1986