TY - JOUR
T1 - Charnley wear model for validation of hip simulatorsball diameter versus polytetrafluoroethylene and polyethylene wear
AU - Clarke, I. C.
AU - Good, V.
AU - Anissian, L.
AU - Gustafson, A.
N1 - Abstract Wear rates of polytetrafluoroethylene (PTFE) and polyethylene cups were compared in 9-channel and 12-channel simulators, using serum lubrication and gravimetric techniques for wear assessment. Cobalt-chromium (CoCr) and alumina ceramic femoral heads in 22-42 mm diameter range were used to validate simulator wear rates against clinical data.
PY - 1997/1
Y1 - 1997/1
N2 - Wear rates of polytetrafluoroethylene (PTFE) and polyethylene cups were compared in 9-channel and 12-channel simulators, using serum lubrication and gravimetric techniques for wear assessment. Cobalt-chromium (CoCr) and alumina ceramic femoral heads in 22-42 mm diameter range were used to validate simulator wear rates against clinical data. This was also the first study of three femoral head sizes evaluated concurrently in a simulator (with three replicate specimens) and also the first report in which any wear experiments were repeated. Fluid absorption artefacts were within +1 per cent of wear magnitude for PTFE and ±8 per cent for polyethylene and were corrected for. Wear rates were linear as a function of test duration. Precision within each set of three cups was within ±6 per cent. The wear rates from experiments repeated over 15 months were reproducible to within ±24 per cent. However, the magnitudes of the simulator wear rates were not clinically accurate, the PTFE wear rates (2843mm3/106 cycles; 22mm diameter) were over three times higher than in vivo, the polyethylene 30 to 50 per cent on the low side (23 mm3/106 cycles; 22 mm diameter). Volumetric wear rate increased with respect to size of femoral head and a linearly increasing relationship of 7-8 per cent/mm was evident with respect to femoral head diameter for both PTFE and polyethylene. These data compared well with the clinical data.
AB - Wear rates of polytetrafluoroethylene (PTFE) and polyethylene cups were compared in 9-channel and 12-channel simulators, using serum lubrication and gravimetric techniques for wear assessment. Cobalt-chromium (CoCr) and alumina ceramic femoral heads in 22-42 mm diameter range were used to validate simulator wear rates against clinical data. This was also the first study of three femoral head sizes evaluated concurrently in a simulator (with three replicate specimens) and also the first report in which any wear experiments were repeated. Fluid absorption artefacts were within +1 per cent of wear magnitude for PTFE and ±8 per cent for polyethylene and were corrected for. Wear rates were linear as a function of test duration. Precision within each set of three cups was within ±6 per cent. The wear rates from experiments repeated over 15 months were reproducible to within ±24 per cent. However, the magnitudes of the simulator wear rates were not clinically accurate, the PTFE wear rates (2843mm3/106 cycles; 22mm diameter) were over three times higher than in vivo, the polyethylene 30 to 50 per cent on the low side (23 mm3/106 cycles; 22 mm diameter). Volumetric wear rate increased with respect to size of femoral head and a linearly increasing relationship of 7-8 per cent/mm was evident with respect to femoral head diameter for both PTFE and polyethylene. These data compared well with the clinical data.
KW - Cup
KW - Head diameter
KW - Hip simulator
KW - Polyethylene
KW - Polytetrafluoroethylene
KW - Wear
UR - https://www.scopus.com/pages/publications/0030625737
UR - https://www.scopus.com/pages/publications/0030625737#tab=citedBy
U2 - 10.1243/0954411971534656
DO - 10.1243/0954411971534656
M3 - Article
C2 - 9141888
SN - 0954-4119
VL - 211
SP - 25
EP - 36
JO - Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
JF - Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
IS - 1