Case report: Complex arrhythmia induced by noncompetitive atrial pacing algorithm of DDDR pacemaker

S. Serge Barold, Paul A. Levine, Dale M. Isaeff, Robert Betzold

Research output: Contribution to journalArticlepeer-review

Abstract

This report describes the occurrence of a repetitive nonreentrant ventriculoatrial (VA) synchronous rhythm precipitated by the noncompetitive atrial pacing algorithm of a Medtronic DDDR pacemaker. This algorithm delivers an atrial stimulus 300 ms after the detection of an atrial signal in the postventricular atrial refractory period of the pacemaker. In our patient, the atrial stimulus released by the algorithm was ineffectual because it encountered prolonged refractoriness of the atrial myocardium. This situation produced a repetitive nonreentrant VA1 synchronous rhythm in the setting of retrograde VA conduction.