T1 - Alumina hip joints characterized by run-in wear and steady-state wear to 14 million cycles in hip-simulator model
AU - Oonishi, Hironobu
AU - Clarke, Ian C.
AU - Good, Victoria
AU - Amino, Hirokazu
AU - Ueno, Masuru
N1 - Corresponding Author E-mail address: [email protected] Peterson Tribology Laboratory, Department of Orthopaedics, LLUMC, 11406 Loma Linda Drive, Suite 606, Loma Linda, California 92354 Peterson Tribology Laboratory, Department of Orthopaedics, LLUMC, 11406 Loma Linda Drive, Suite 606, Loma Linda, California 92354 Search for more papers by this author First published: The biphasic wear performance (run‐in; steady‐state phase) of 28‐mm alumina‐alumina hip implants was studied by hip simulator methods using bovine serum as the lubricant.
PY - 2004/9/15
Y1 - 2004/9/15
N2 - The biphasic wear performance (run-in; steady-state phase) of 28-mm alumina-alumina hip implants was studied by hip simulator methods using bovine serum as the lubricant. The Biolox™ implants were run to 5.7 million cycles and Bioceram™ implants to 14.4 million cycles (Mc). Wear with all-alumina total hip replacements (THR) first showed a high wear rate of the order 1.2 mm3/Mc, lasting approximately 0.17 Mc. Overall to 0.7 Mc, the run-in phase appeared curvilinear but could be described by a linear phase averaging 0.3 mm3/Mc. From 0.7 to 1 Mc duration, the wear trend transitioned into a steady-state phase. Wear rates from 1 to 14 Mc were of the order 0.02 mm3/Mc. Surface contamination from the serum lubricant resulted in cyclic weight fluctuations of the order 0.2 mg. The transition from average run-in to steady-state phase represented a wear reduction of 13-fold. Comparing the steady-state wear value to that in standard 28-mm UHMWPE CUPS approaching 75 mm3/year, there was clearly a three-orders-of- magnitude wear superiority in favor of ceramic cups. © 2004 Wiley Periodicals, Inc.
AB - The biphasic wear performance (run-in; steady-state phase) of 28-mm alumina-alumina hip implants was studied by hip simulator methods using bovine serum as the lubricant. The Biolox™ implants were run to 5.7 million cycles and Bioceram™ implants to 14.4 million cycles (Mc). Wear with all-alumina total hip replacements (THR) first showed a high wear rate of the order 1.2 mm3/Mc, lasting approximately 0.17 Mc. Overall to 0.7 Mc, the run-in phase appeared curvilinear but could be described by a linear phase averaging 0.3 mm3/Mc. From 0.7 to 1 Mc duration, the wear trend transitioned into a steady-state phase. Wear rates from 1 to 14 Mc were of the order 0.02 mm3/Mc. Surface contamination from the serum lubricant resulted in cyclic weight fluctuations of the order 0.2 mg. The transition from average run-in to steady-state phase represented a wear reduction of 13-fold. Comparing the steady-state wear value to that in standard 28-mm UHMWPE CUPS approaching 75 mm3/year, there was clearly a three-orders-of- magnitude wear superiority in favor of ceramic cups. © 2004 Wiley Periodicals, Inc.
KW - Alumina
KW - Ceramic
KW - Hip arthroplasty
KW - Simulator
KW - Wear
UR - https://www.scopus.com/pages/publications/4344667565
UR - https://www.scopus.com/pages/publications/4344667565#tab=citedBy
UR - https://www.mendeley.com/catalogue/443c5aa8-4796-30b4-8def-5e2656a101e4/
U2 - 10.1002/jbm.a.30021
DO - 10.1002/jbm.a.30021
M3 - Article
C2 - 15307156
SN - 0021-9304
VL - 70
SP - 523
EP - 532
JO - Journal of Biomedical Materials Research - Part A
JF - Journal of Biomedical Materials Research - Part A
IS - 4
ER -