Alumina hip joints characterized by run-in wear and steady-state wear to 14 million cycles in hip-simulator model

  • Hironobu Oonishi
  • , Ian C. Clarke
  • , Victoria Good
  • , Hirokazu Amino
  • , Masuru Ueno

Research output: Contribution to journal › Article › peer-review

Abstract

The biphasic wear performance (run-in; steady-state phase) of 28-mm alumina-alumina hip implants was studied by hip simulator methods using bovine serum as the lubricant. The Biolox™ implants were run to 5.7 million cycles and Bioceram™ implants to 14.4 million cycles (Mc). Wear with all-alumina total hip replacements (THR) first showed a high wear rate of the order 1.2 mm3/Mc, lasting approximately 0.17 Mc. Overall to 0.7 Mc, the run-in phase appeared curvilinear but could be described by a linear phase averaging 0.3 mm3/Mc. From 0.7 to 1 Mc duration, the wear trend transitioned into a steady-state phase. Wear rates from 1 to 14 Mc were of the order 0.02 mm3/Mc. Surface contamination from the serum lubricant resulted in cyclic weight fluctuations of the order 0.2 mg. The transition from average run-in to steady-state phase represented a wear reduction of 13-fold. Comparing the steady-state wear value to that in standard 28-mm UHMWPE CUPS approaching 75 mm3/year, there was clearly a three-orders-of- magnitude wear superiority in favor of ceramic cups. © 2004 Wiley Periodicals, Inc.
Original languageEnglish
Pages (from-to)523-532
Number of pages10
JournalJournal of Biomedical Materials Research - Part A
Volume70
Issue number4
DOIs
StatePublished - Sep 15 2004

ASJC Scopus Subject Areas

  • Ceramics and Composites
  • Biomaterials
  • Biomedical Engineering
  • Metals and Alloys

Keywords

  • Alumina
  • Ceramic
  • Hip arthroplasty
  • Simulator
  • Wear

Cite this